A Study on the Image Uniformity According to Center-line Velocity in Digital Radiation System

Authors

  • Byeong-Jin Lee master, Dept. of Radiological Science, Hanseo University, Seosan-si, 31962, Korea Author
  • Cheong-Hwan Lim Professor, Dept. of Radiological Science, Hanseo University, Seosan-si, 31962, Korea Author
  • Seung-Hun Jeong Doctor, Dept. of Radiology, Seosan Jung-Ang Hospital, Seosan-si, 32010, Korea Author

DOI:

https://doi.org/10.61841/d587c754

Keywords:

Digital Radiation System, Central Line, Uniformity, Signal to Noise Ratio, Intensity Distribution

Abstract

Background/Objectives: Generally speaking, it is the case that there is an increase in diagnostic radiators are being examined without aligning the center-line, due to its movement constraint. Therefore, we would like to compare the uniformity of images according to the center-line velocity in the CR System and DR System.

Methods/Statistical Analysis: The methods used in this study are to measure the uniformity of the images, as the images were acquired by preparing a 7x7x7(cm) hexagonal phantom with 3mm thick acrylic and adding water. As noted in the CR and DR systems, SID was measured at 150cm and 180cm, and the measurement condition was shown to be 50kVp and 4mAs for each one.

Findings: The findings noted that the doses were shown to have a higher cathode value than the center value according to the Heel Effect of the radiation equipment, and were measured lower in the Anode. Although the SNR showed little change in location in the CR System, the difference between the Cathode and Anode in the DR system was more than 8%. In this context, the intensity distribution in this case was shown to show a large variation of the cathode and anode in the CR System with decreasing in the uniformity of the images. Also, it is noted that in the DR system, the uniformity of the image decreased due to a large variation in the anode.

Improvements/Applications: In this case, the discordance of the center-line in the DR system was shown to have a large impact on the image and reduced uniformity. Overall in the process of the radiation test, it was analyzed that the degree of uniformity is excellent when the detector and center-line are accorded.

 

Downloads

Download data is not yet available.

References

[1] Park YK, Jung SE. CT radiation dose and radiation reduction strategies. Journal Korean Medical Association. 2011 December;54(12):1262-8.

[2] http://dx.doi.org/10.5124/jkma.2011.54.12.1262

[3] Jang HY, Choi JI, Jung SE, Rha SE, Oh SN, Lee YJ, et al. Radiation Dose and Imaging Quality of Abdominal Computed Tomography before and after Scan Protocol Adjustment: Single Institutio n Experience in Three Years. Journal Korean Society Radiology 2014 June;71(6):278-87.

a. http://dx.doi.org/10.3348/jksr.2014.71.6.278

[4] Lee W. Current status of medical radiation exposure and regulation efforts. Journal Korean Medic al Association. 2011December;54(12): 1248-52.

a. http://dx.doi.org/10.5124/jkma.2011.54.12.1248

[5] Joo YC, Lim CH, You IG, Jung HR, Lee SH. Adequacy of Source to

a. Image Receptor Distance with Chest Postero-Anterior Projection in DigitalRadiology Syst em. Journal of Radiological Science and Technology, 2016 June;39(2):35-142.

[6] https://doi.org/10.17946/JRST.2016.39.2.01

[7] Min JW, Kim JM, Jeong HW. Artifacts in Digital Radiography. Journal

a. of Radiological Science and Technology. 2015 December;38(4):375-81.

b. https://doi.org/10.17946/JRST.2015.38.4.06

[8] Kim KW, Park JH, Yoon SH. Medical Ethics in Radiology. Journal Korean Society Radiology. 20 10 Apr; 62(4):311-7.

a. https://synapse.koreamed.org/search.php?where=aview&id=10.3348/jksr.2010.62.4.311&cod e=2016JKSR&vmode=PUBREADER

[9] Kim G, Lee R. Effect of Target Angle and Thickness on the Heel Effectand X-ray Intensity Chara cteristics for 70 kV X-ray Tube Target PROGRESS in MEDICAL PHYSICS. 2016 December;27( 4):272-6.

[10] https://doi.org/10.14316/pmp.2016.27.4.272

[11] Lee DY, Lee JS. Evaluation of the Space Scattered Dose According to the Position of the Radi ation Workers in Mammography Room. Journal of Radiological Science and Technology. 2016 Se ptember; 39(3):297-303.

[12] https://doi.org/10.17946/JRST.2016.39.3.01

[13] Choi SS, Lim CH, Jung SH. A study on the Using of Automatic Exposure Control in the Chest Radiography. Journal of Radiological Science and Technology. 2019 January; 42(1):19-24.

a. https://doi.org/10.17946/JRST.2019.42.1.19

[14] An BJ. A comparative study for resolution and density of chest imaging using film/screen CR and DR. Journal of Korean Society of Radiology. 2010 March; 4(1):25-30.

a. http://dx.doi.org/10.7742/jksr.2010.4.1.025

Downloads

Published

30.09.2020

How to Cite

Lee, B.-J., Lim, C.-H., & Jeong, S.-H. (2020). A Study on the Image Uniformity According to Center-line Velocity in Digital Radiation System. International Journal of Psychosocial Rehabilitation, 24(7), 2304-2311. https://doi.org/10.61841/d587c754