Examination of Signal Intensity According to the Alteration of Contrast Media Concentration during Computed Tomography

Authors

  • Sung-Hun Jeong Public Health of Doctor, SeoSan JungAng General Hospital, Seosan-si, 31962, Korea Author
  • Cheong-Hwan Lim Professor, Dept. of Radiological Science, Hanseo University, Seosan-si, 31962, Korea Author

DOI:

https://doi.org/10.61841/74xrte58

Abstract

Background / Objectives: Generally speaking, it is shown that many medical institutions use contrast agents for both children and adults without distinguishing the concentration of the contrast agent, and many healthcare providers use the same contrast agent regardless of the examination site as identified on the patient. This is a factor that can increase the side effects by proceeding the test without considering the sensitivity, the concentration of the contrast agent, the injection speed, etc., depending on the patient's constitution that causes the side effects of the contrast agent. For this reason, the purpose of this study is to obtain the image according to the concentration of CT contrast agent and test condition, to identify the change in signal intensity, and to determine the appropriate test condition according to the contrast agent concentration and to provide the study results thereof as the basic material in the study on the image concentration according to the change of the test condition.

Methods / Statistical analysis: The equipment used in this study was MDCT and contrast agents were used at concentrations of 250, 300, 320, 350, 370, and 400mgI/ml. In this case, the contrast agent was measured after placing the Conical Tube on the table of MDCT without dilution. Here, the measurement conditions were measured at 37, 75, 100, 150, 320mAs for tube current mAs and 80, 100, 120, 135kVp.

Findings: The findings revealed that as the tube voltage increased at tube currents of 75mAs, 100mAs, 150mAs, and 320mAs, the signal strength of the image was noted to be lowered in this case. In addition, the signal strength of the image was lowered as the tube voltage increased according to the contrast agent concentration.

Improvements / Applications: It is considered that further studies will be needed for the test protocols that can lead to the improvement of the image quality safely and steadily, as the study of the appropriate contrast agent concentration and the test conditions for each test is continued to help improve patient outcomes.

Keywords- Iodide contrast agent, Signal Intensity, Contrast agent side effects, Contrast Concentration, Computed Tomography

Downloads

Download data is not yet available.

References

[1] Ko IH, Kwon DC, Kim KK, Kim KH, Kim MK, Kim MC. TEXTBOOK of Computed Tomograp hy: Chung-Ku publishing; 2017. p. 127-52.

a. http://www.yes24.com/Cooperate/Naver/welcomeNaver.aspx?pageNo=1&goodsNo=7610528 5

[2] Lee SJ, Park Y, Jeong DH, Ock SH, Hwang HJ, Hwang HS. A Study on the usefulness of Diluti on of Contrast Media in CT Examinations. Journal of Korean society of computed tomographic te chnology. 2011 October;13(2):161-7.

a. http://scholar.dkyobobook.co.kr.library.hanseo.ac.kr:8000/searchDetail.laf?barcode=4010026 788675#

[3] Yang WT, Shin SH. Research on contrast media injection method by Heart rate for diagnostic are a extension when do Cardiac CTA. Journal of Korean society of computed tomographic technolog

y. 2008 April;10(1):89-95.

a. http://scholar.dkyobobook.co.kr.library.hanseo.ac.kr:8000/searchDetail.laf?barcode=4010026 526893

[4] Kim MJ, Choi JS, Moon JW, Bum HN, Jeon JS, Choi NG. Study on contrast warmer temperature and heating time for appropriate contrast media during the CT examination. Journal of Korean so ciety of computed tomographic technology. 2016 March;18(1):95-101.

a. http://scholar.dkyobobook.co.kr.library.hanseo.ac.kr:8000/searchDetail.laf?barcode=4010026

148523

[5] Yoo SJ, Seon JR, Oh SJ, Rhim JD, Bae SH. Contrast media reactions: Cause analysis during CT examination. korea safety management & science. 2014 November; 2014(11):223~9.

a. http://db.koreascholar.com.library.hanseo.ac.kr:8000/Article?code=320573

[6] Kim HH, Choi JY, Oh MK, Kim EY, Ghim JR, Choi SJ, et al. Retrospective Analysis of Advers e Reactions to Iodinated Contrast Media in Korean. Translational and Clinical Pharmacology. 201 2 december;20(2):165-74.

a. http://www.riss.kr.library.hanseo.ac.kr:8000/search/detail/DetailView.do?p_mat_type=1a0202e37d52c7 2d&control_no=360877cca93735ce47de9c1710b0298d

[7] Lee KO. Study for Correlation between body composition and HU value of Portal Vein for Amou nts of Contrast Medium in Abdomen CT Examination[Master]. [Jeon ju]: Dept. of Radiological Sc ience Graduate School of International University of Korea; 2011. 51~5 p.

a. http://www.riss.kr.library.hanseo.ac.kr:8000/search/detail/DetailView.do?p_mat_type=be54d9b8bc7cdb 09&control_no=6e730ba0bcacb8cbffe0bdc3ef48d419

[8] Jang YS, Lee GW, Ku YSu, Kim DH. The usefulness of pretreatment for adverse reactions to contrast media high‐risk group. Journal of Korean society of computed tomographic technology. 2013 October; 15(1):201~8.

a. http://scholar.dkyobobook.co.kr.library.hanseo.ac.kr:8000/searchDetail.laf?barcode=4010026220357#

[9] Kwon KS, Jeong JS. An alts is of Adverse Reactions to Computed Tomography Contrast Medium. Journal of korean biological nursing science. 2004 december;6(2) 57-68.

a. http://scholar.dkyobobook.co.kr.library.hanseo.ac.kr:8000/searchDetail.laf?barcode=4050025673217

[10] Korea Consumer Resources. Investigation on safety status of contrast agent. 2017;10-5

a. https://www.kca.go.kr/home/sub.do?menukey=4062&mode=view&no=1002617385&page=11

[11] Jung SC. Comparison of the Use of Contrast Media with Different Iodine concentrations for Multi detector-Row CT of the Kidney[master]. [seoul]: A thesis submitted in partial fulfillment of The r equirement for the degree of Master in medicine.(Radiology) in the seoul National University, seo ul. 2008. 8~15p.

a. http://www.riss.kr.library.hanseo.ac.kr:8000/search/detail/DetailView.do?p_mat_type=be54d9b8bc7cdb 09&control_no=3fd95f8398e2dc74ffe0bdc3ef48d419

[12] Lee SW, Kim HJ, Kim TH, Jo SJ. Estimation and Application of HU Values for Various Material s as a Function of Physical Factor. Korean Journal of Medical Physics. 2009 September;20(3): 14 5-51.

a. http://www.ndsl.kr.library.hanseo.ac.kr:8000/ndsl/search/detail/article/articleSearchResultDetail.do?cn=

JAKO200914064135825&SITE=CLICK

[13] Kang HR, Lee YJ. Risk Factors for Adverse Reactions to Iodinated Contrast Media in Computed Tomography. Korean Journal Clinical Pharm. 2016 February; 26(1): 40-3.

a. http://www.riss.kr.library.hanseo.ac.kr:8000/search/detail/DetailView.do?p_mat_type=1a0202e37d52c7 2d&control_no=453a04cbb38eafd7b7998d826d417196

[14] Yu SM. Comparison of Iodine Constrast Media for Adverse Reactions of Contrast Media in Com puted Tomography[master]. [seoul]: Major in Biomedical Science, Department of Health and Conv ergence Science The Graduate School, Korea University. 2018. 6~10p.

a. http://www.riss.kr.library.hanseo.ac.kr:8000/search/detail/DetailView.do?p_mat_type=be54d9b8bc7cdb 09&control_no=83aa7f2b77973bc2ffe0bdc3ef48d419

[15] Kim YK, Kim MS, Lee SY, Lee JM, Lee HS, Jeong JS et al. Comparison of the contrast effect o f the amount and iodine concentration of CT contrast agent. Korean Society of Computed Tomogr aphic Technology. 2014 October; 16(2):183~4.

a. http://scholar.dkyobobook.co.kr.library.hanseo.ac.kr:8000/searchDetail.laf?barcode=4010026220478

[16] Korean statistical information service. Status of Special Medical Equipment by Region.

a. http://kosis.kr/statHtml/statHtml.do?orgId=354&tblId=DT_HIRA4I

[17] Health Insurance Review and Assessment Service, Health care big data open system, CT Status b y Nursing Institution Location.

a. http://opendata.hira.or.kr/op/opc/olapMfrnIntrsDiagBhvInfo.do

[18] Jung KK, Cho PK. Evaluation of the Low Tube Voltage in the Computed Tomography Scan Tech nique using a Low Concentration Contrast Agent. Journal of Radiological Science and Technology

. 2016 March; 39(1): 71~4.

a. http://scholar.dkyobobook.co.kr.library.hanseo.ac.kr:8000/searchDetail.laf?barcode=4010024834028

Downloads

Published

30.09.2020

How to Cite

Jeong, S.-H., & Lim, C.-H. (2020). Examination of Signal Intensity According to the Alteration of Contrast Media Concentration during Computed Tomography. International Journal of Psychosocial Rehabilitation, 24(7), 2040-2049. https://doi.org/10.61841/74xrte58