Study on the Dose and Exposure Characteristics of the Automatic Optimization Parameter Mode in Mammography

Authors

  • Young-Ju Moon Master’s student, Dept. of Radiological Science, The Graduate School of Hanseo University, 46, Hanseo 1-ro, Haemi-myun, Seosan-si, Chungcheognam-do, Province, 31962, Republic of Korea Author
  • Hong-Ryang Jung Professor, Dept. of Radiological Science, The Graduate School of Hanseo University, 46, Hanseo 1-ro, Haemi-myun, Seosan-si, Chungcheognam-do, Province, 31962, Republic of Korea Author
  • Woo-Taek Lim Ph.D, Dept. of Radiology, Konkuk University Medical Center, 120-1, Neungdong-ro, Gwangjin-gu, Seoul, 05030, Republic of Korea Author

DOI:

https://doi.org/10.61841/mz8x9t78

Keywords:

digital mammography, automatic optimization parameter, image quality, average glandular dose, Entrance skin exposure, , polymethylmethacrylate.

Abstract

Background/Objectives: The purpose of study was to provide useful information on the choice of the AOP mode of units that can improve the quality of breast images and reduce exposure dose by analyzing the dose and exposure characteristics of the AOP mode used in the mammography unit of the digital system.

Methods/Statistical analysis: The equipment used in the experiment is GE’s Senographe Pristna and Hologic's Selenia Dimensions. The phantom used Polymethylmethacrylate phantoms and ACR phantom for the evaluation each according to the automatic exposure control of parameters mode for each equipment to evaluate the average dose. Statistical analysis was performed using one-way ANOVA, and post-hoc was performed using the Tukey method. The statistically significant range was set to (p<0.05).

Findings: In the dose evaluation, GE’s Senographe Pristna showed that AGD and ESE were lower in the Dose- mode and, while Hologic's Selenia Dimensions showed that Auto Time mode was lower in AGD and ESE than the Auto Filter mode in dose evaluation. When compared on the whole, both AGD and ESE showed the lowest dose in the Dose-mode, but no statistical difference was found (p>0.05). In the image evaluation, all the scene modes used in the experiment showed different high and low evaluation scores for each item in the evaluation criteria range, but the Dose-mode is believed to provide high-quality images in diagnosing diseases of the breast, as well as mitigating exposure dose since the Dose-mode was the highest in the calcification detection that is considered crucial in mammography.

Improvements/Applications: These results showed that AOP modes applied to mammography units in digital systems can improve the quality of breast images and reduce exposure dose. It is also posited that the results of this study will be used as a baseline data for establishing optimized guidelines for image quality improvement and dose reduction in future mammography examinations.

Downloads

Download data is not yet available.

References

[1] Jang S-B, Yoo D-Y. Effects of Chelidonium majus extract on apoptosis induction of MDA-MB-231 human breast cancer cells. The Journal of Korean Obstetrics and Gynecology. 2015;28(2):15-25.

[2] Han G-j, Lee Ar, Seong S, Kim S-s. A Case Study of an Advanced Breast Cancer Patient Treated with a Combination of Traditional Korean Medicine and Chemotherapy. J Int Korean Med. 2018;39(2):268-76.

[3] Kim SJ, Lee JM, Min HY, Min HY, Kim SJ, Lee JM, et al. Relationship between Knowledge, Attitude and Practice of Breast-Self Examination among Middle and High School Girls. Child Health Nursing Research. 2017;23(2):147-57.

[4] Noh DY, Kim JS, Youn YK, Oh SK, Choe KJ. Changes in the Clinical Features of and the Treatment for Breast Cancer. Journal of the Korean Surgical Society. 1998;54(4):464-73.

[5] Kim H-M, Chon KS. Changes in Spatial Resolution at Position of the Detector in Digital Mammography System. Journal of the Korean Society of Radiology. 2016;10(3):215-22.

[6] Rodríguez-Ruiz A, Krupinski E, Mordang J-J, Schilling K, Heywang-Köbrunner SH, Sechopoulos I, et al. Detection of breast cancer with mammography: effect of an artificial intelligence support system. Radiology. 2019;290(2):305-14.

[7] Lee EH, Kim KW, Kim YJ, Shin D-R, Park YM, Lim HS, et al. Performance of screening mammography: a report of the alliance for breast cancer screening in Korea. Korean journal of radiology. 2016;17(4):489- 96.

[8] Chen B, Wang Y, Sun X, Guo W, Zhao M, Cui G, et al. Analysis of patient dose in full field digital mammography. European journal of radiology. 2012;81(5):868-72.

[9] Byung Sam Kang, Hwa Yeon Yeo, Dong KR. Image Comparison Analysis of Film-Screen Methods Which Use Mammo Phantom and DR Methods. J of Advanced Engineering and Technology. 2013;6(1):43-7.

[10] Brooks H. The relationship between science and technology. Research policy. 1994;23(5):477-86.

[11] Huda W, Sajewicz AM, Ogden KM, Scalzetti EM, Dance DR. How good is the ACR accreditation phantom for assessing image quality in digital mammography? Academic radiology. 2002;9(7):764-72.

[12] Shim BY. Cancer Screening Guidelines in Korea. Korean J Med. 2016;90(3):224-30.

[13] Bassett L, Hirbawi I, DeBruhl N, Hayes M. Mammographic positioning: evaluation from the view box. Radiology. 1993;188(3):803-6.

[14] Lee JA, Yu JH, Song Y-M. Management of long-term breast cancer survivors in Korea. Journal of the Korean Medical Association. 2016;59(4):266-75.

[15] Kamal I, Chelliah KK, Mustafa N. Estimates of Average Glandular Dose with Auto-Modes of X-Ray Exposures in Digital Breast Tomosynthesis. Sultan Qaboos University Medical Journal. 2015;22(2617):1- 5.

[16] Ko M-S, Kim HH, Cha JH, Shin HJ, Kim JH, Kim MJ. Dose reduction in automatic optimization parameter of full field digital mammography: Breast phantom study. Journal of breast cancer. 2013;16(1):90-6.

Downloads

Published

30.09.2020

How to Cite

Moon, Y.-J., Jung, H.-R., & Lim, W.-T. (2020). Study on the Dose and Exposure Characteristics of the Automatic Optimization Parameter Mode in Mammography. International Journal of Psychosocial Rehabilitation, 24(7), 1839-1845. https://doi.org/10.61841/mz8x9t78